AlazarAbortAsyncRead#

Function Syntax#

RETURN_CODE AlazarAbortAsyncRead(HANDLE handle)#

Aborts a dual-port acquisition, and any in-process DMA transfers.

Remark

If you have started an acquisition and/or posted DMA buffers to a board, you must call AlazarAbortAsyncRead() before your application exits. If you do not, when your program exists, Microsoft Windows may stop with a blue screen error number 0x000000CB (DRIVER_LEFT_LOCKED_PAGES_IN_PROCESS). Linux may leak the memory used by the DMA buffers.

Note

This function is part of the dual-port API. It should be used only in this context. To abort single-port acquisitions using, see AlazarAbortCapture().

Parameters
  • handle – [in] Handle to board

Returns

ApiSuccess upon success, or an error code. See RETURN_CODE for more detailed information.

LabVIEW Block Diagram#

../_images/AlazarAbortAsyncRead.vi.png